BA229R/09/en/07.0771025405Release 01.02Operating manualThermophant® T TTR31, TTR35Temperature switch
Operation TTR31, TTR3510 Endress+Hauser4. 2 DC voltage version with valve connectorP01-PTx3xxxx-04-xx-xx-xx-003Fig. 6: Thermophant® T with M 16x1.5 o
TTR31, TTR35 OperationEndress+Hauser 115.1.1 Navigating in the operating menuT09-TTR31xxx-19-xx-xx-xx-002Fig. 8: Navigating in the operating menuA Fun
Operation TTR31, TTR3512 Endress+Hauser5.1.2 Structure of the operating menu for 1x or 2x switch outputsT09-TTR31xxx-19-xx-xx-xx-003Fig. 9: Operating
TTR31, TTR35 OperationEndress+Hauser 135.1.3 Structure of the operating menu for 1x switch output or1x analog output (4 to 20 mA)At devices with analo
Operation TTR31, TTR3514 Endress+Hauser5.1.4 Basic settings!Note! The function group 4-20 is only available if the 4 to 20 mA analog output (0>.,)
TTR31, TTR35 OperationEndress+Hauser 155.1.5 Settings for output - 2x switch output• Hysteresis functionThe hysteresis function enables two-point cont
Operation TTR31, TTR3516 Endress+HauserT09-TTR31xxx-05-xx-xx-xx-001Fig. 11: Switch point functionsa: Hysteresis-functionb: Window-function m Window -
TTR31, TTR35 OperationEndress+Hauser 175.1.6 Settings for output - 1x switch output and 1x analog output (4 to 20 mA)Output 1.Output 2, as optio
Operation TTR31, TTR3518 Endress+HauserOutput 1.Output 2, as option.Switch-back point value,M, Switch-back point -50 to 149 °C(-58 to 300
TTR31, TTR35 OperationEndress+Hauser 19!Note! The function group 4-20 is only available if the 4 to 20 mA analog output (0>.,) is selected in the f
Brief overviewUsing the following short form instructions you can commission your system easily and swiftly:Safety instructions → Page 4ÆInstallation
Operation TTR31, TTR3520 Endress+HauserService functions.(if output 2 available)Simulationoutput 1 or 266/M1(if analog output avail
TTR31, TTR35 OperationEndress+Hauser 215. 2 Operation with PCThe device can be configured with the configuration software ReadWin® 2000 or FieldCare.
Maintenance TTR31, TTR3522 Endress+Hauser5.2.2 Hints for the configuration with Readwin® 2000Comprehensive information on the ReadWin ® 2000 configura
TTR31, TTR35 AccessoriesEndress+Hauser 237 AccessoriesIf ordering accessories, please specify the serial number of the unit! All dimensions in the dra
Accessories TTR31, TTR3524 Endress+Hauser!Note! We recommend to change the O-ring in the same time frame as of all other sealings in yourprocess.7.1.2
TTR31, TTR35 AccessoriesEndress+Hauser 257.1.3 O-ring for adapter replacementO-ring 15.54 x 2.62 mm (0.61 x 0.1 in), EPDM 70 Shore FDA, order no. 5100
Accessories TTR31, TTR3526 Endress+Hauser7.2 Welding bosses and coupling 7.2.1 Welding boss with sealing taper7.2.2 Collar welding boss7.2.3 Weldin
TTR31, TTR35 AccessoriesEndress+Hauser 277.2.4 Coupling with sealing taper7.3 Electrical connection7.3.1 Plug-in jack; connecting cableMoveable coupli
Accessories TTR31, TTR3528 Endress+Hauser7.4 Configuration kitPVC cable, 4 x 0.34 mm2 (22 AWG) with M12x1 coupling, with LED, elbowed, 316L screw plu
TTR31, TTR35 Trouble-shootingEndress+Hauser 298 Trouble-shooting8. 1 Errors and warningsIf an error in the device occurs, the colour of the status LE
Endress+Hauser 3TTR31, TTR35Table of contents1 Safety instructions . . . . . . . . . . 41. 1 Designated use . . . . . . . . . . . . . . . . . .
Trouble-shooting TTR31, TTR3530 Endress+Hauser8. 2 RepairA repair is not planned.8. 3 DisposalPlease pay particular attention to the local disposal
TTR31, TTR35 The most important technical dataEndress+Hauser 318. 4 Software history and compatibility overviewThe release number on the nameplate an
The most important technical data TTR31, TTR3532 Endress+Hauser9. 2 OutputSwitching capacity• Switch status ON: Ia ≤ 250 mA• Switch status OFF: Ia ≤
TTR31, TTR35 The most important technical dataEndress+Hauser 33Process pressure limitsMaximum permitted process pressure depending on the insertion le
The most important technical data TTR31, TTR3534 Endress+HauserPermitted flow velocity depending on the insertion length.a0008065-enFig. 15: Permitted
www.endress.com/worldwide BA229R/09/en/07.07 71025405 CCS/FM6.0+SGML
Safety instructions TTR31, TTR354 Endress+Hauser1 Safety instructions1. 1 Designated useThe Thermophant® T is a temperature switch for monitoring, di
TTR31, TTR35 Device identificationEndress+Hauser 52 Device identification2. 1 NameplateTo identify your device, compare the complete order code and t
Installation TTR31, TTR356 Endress+HauserUL listed for Canada and USAThe device was examined by Underwriters Laboratories Inc. (UL) in accordance with
TTR31, TTR35 InstallationEndress+Hauser 73. 3 Process connectionT09-TTR31xxx-17-xx-xx-en-000Fig. 3: Process connection of the devices"Caution! T
Installation TTR31, TTR358 Endress+Hauser3. 4 Installation instructionsT09-TTR31xxx-11-00-xx-xx-000Fig. 4: Possible installation options for temperat
TTR31, TTR35 WiringEndress+Hauser 94 Wiring4. 1 DC voltage version with M12x1 connectorP01-PTx3xxxx-04-xx-xx-xx-002Fig. 5: Thermophant® T with M12x1