Velocity Edge DualX Manual de Serviço Página 340

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 400
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 339
316 MARK III mechanical/electrical troubleshooting
Troubleshooting comm., mech., and elec. Reference, Installation, and Operations Manual
August 2014 3-9000-743 Rev U
communicating with meter but all
chords display failures
Verify that the resistance of transducers is within Specification (1-2 Ω).
Check the Acquisition board (see Appendix E).
Check the interconnect cables between the base assembly and the head assembly
(
see Appendix E).
chord is not indicating
Check the resistance of the failed transducer.
If Chord A is not indicating, change the transducer cables from Chord D to chord A.
If Chord D then fails, the transducers are bad on Chord A (
see Appendix E).
The same test procedure can be accomplished by swapping Chords B and C if a
chord fails in either chord (
see Appendix E).
Note that the outside chord cables cannot be exchanged with inner
chord cables.
Table D-1 Mark III Mechanical/Electrical Troubleshooting
Problem Solution(s)
Vista de página 339
1 2 ... 335 336 337 338 339 340 341 342 343 344 345 ... 399 400

Comentários a estes Manuais

Sem comentários